A flexible, dependable and cost-efficient module blood pumping system.
Features
Modular design makes a custom system affordable
Four or five pump base
Roller or centrifugal pumps
Arterial and cardioplegia monitors for temperature, pressures and timers
Battery module
Electronic venous occluder
Pulse module
Communications module for data acquisition
It is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion and cardiopulmonary bypass procedures only
Especificaciones
Electrical Power Requirements
115 V Bases:
Voltage - ~103.5-126.5 V
Frequency - 50/60 Hz
Circuit Breaker Protection - 16 amps
Fully Loaded System Current Rating - 4 pump base 13 amps, 5 pump base 15 amps
220-240 V Bases:
Voltage - ~1198-264 V
Frequency - 50/60 Hz
Circuit Breaker Protection - 8 amps
Fully Loaded System Current Rating - 4 pump base 6 amps, 5 pump base 7 amps
100 V Bases:
Voltage - ~90-110 V
Frequency - 50/60 Hz
Circuit Breaker Protection - 16 amps
Fully Loaded System Current Rating - 4 pump base 14 amps, 5 pump base 16 amps
Electrical Power Performance
Leakage Current: <100 microamperes for 115 V and 100 V bases (fully loaded); <200 microamperes for 22-240 V bases (fully loaded). The maximum leakage current for the base without pumps is 45 microamperes for all input voltages
Ground Resistance: 0.1 ohms maximum for all bases.
Emergency Power: 8000 system battery (two 12 volt, sealed gel cell batteries) will power the arterial pump (operating in continuous mode), the monitors, occluder and lamp for 40 minutes; when the arterial pump is operated in pulsatile mode or the cardioplegia pump is also operating on battery power, the battery time available will diminish to 25 minutes. the base is test for high potential: 1200V ~ for 100-1115 VAC units; 1500V ~ for 220-240 VAC units.